German Medical Devices Operator Ordinance (Medizinprodukte-Betreiberverordnung)

A newly revised version of the ordinance has applied since 20 February 2025. For devices under Annex 1, safety checks are required at least every two years; for devices with a measuring function under Annex 2, metrological checks. Operators keep an inventory register and a medical device logbook.

For operators
  • Safety checks at least every two years (§ 12)
  • Metrological checks under Annex 2 and the PTB (national metrology institute) guideline (§ 15)
  • Inventory register for all active non-implantable devices (§ 14)
  • Medical device logbook for devices under Annex 1 and 2 (§ 13)
With vi

vi maintains the inventory register and deadlines from service reports and invoices, reminds you of safety checks (STK) and metrological checks (MTK) and keeps proof ready per device.

Deadlines
STK (safety check)
at least every 2 years
Source
Inspection interval calculator

What applies when.

Adopted, in force, applicable: the milestones of MPBetreibV that matter for service and maintenance.

  1. 14 Feb 2025
    The newly worded Medical Devices Operator Ordinance (MPBetreibV) is issued (Federal Law Gazette 2025 I No. 38). It replaces the 1998 ordinance.
  2. 20 Feb 2025
    The ordinance enters into force.
  3. 1 Aug 2025
    From this day, § 17 (special duties for certain software, including the IT security check) and the ordinance for products under Annex XVI of Regulation (EU) 2017/745 apply.
  4. 4 Nov 2025
    The second amending ordinance is promulgated (Federal Law Gazette 2025 I No. 263, issued on 31 October 2025). It replaces Annex 2 on metrological checks with effect from 20 February 2025.
  5. 1 Jan 2027
    Newly commissioned automated external defibrillators for lay users must document self-test results and transmit them to the operator by remote monitoring (§ 12(2)).

Questions in detail.

Expand a question for the answer, the original wording of the rules and links to the official text.

Who is the operator under the ordinance, and what stays with me if I outsource tasks?

Under § 2(2), the operator is any natural or legal person responsible for running the health care facility in which employees operate or use the device. A health care facility is any facility in which devices are operated or used by medical staff, nursing professionals or other authorized persons, including rehabilitation and nursing facilities. The user is whoever applies a device to a patient (§ 2(3)).

The operator must perform the duties assigned by the ordinance to ensure safe and proper operation and use (§ 3(1)). The operator may assign tasks to service providers but remains responsible. Maintenance, safety checks, metrological checks and IT security checks may only go to persons, businesses or institutions that meet the requirements of § 5 (§ 7(3), § 12(4), § 15(6), § 17(4)).

An exception applies when a supplier (Versorgender) provides devices that remain the supplier's property. Then the supplier has the operator duties and can transfer individual tasks to third parties by contract but must still ensure they are performed properly (§ 3(2)). In daily terms: contracts with service companies should settle the qualification under § 5, the records and the deadline notices. They cannot replace the operator's responsibility.

“The operator of a device is any natural or legal person who is responsible for operating the health care facility in which the device is operated or used by its employees.”
§ 2(2)MPBetreibV (Medical Devices Operator Ordinance)Original text ↗

Unofficial translation

“The operator may only commission persons, businesses or institutions for maintenance who themselves, or whose employees performing the maintenance, meet the requirements of § 5 with regard to the maintenance of the device concerned.”
§ 7(3)MPBetreibV (Medical Devices Operator Ordinance)Original text ↗

Unofficial translation

Which devices need the safety check, and at what interval?

Safety checks (STK) are prescribed for the products in Annex 1. These are non-implantable active products for generating and applying electrical energy to nerves, muscles or the heart, including defibrillators, for measuring in blood vessels, for coagulation or tissue destruction, for introducing substances into the bloodstream with potential pressure buildup, for mechanical ventilation, for diagnosis by nuclear magnetic resonance, for therapy with pressure chambers or hypothermia, as well as infant incubators and external active components of active implants. For other products and objects connected to Annex 1 products, the provisions apply accordingly (§ 12(1) sentence 5).

The checks must be done at least every two years, by the end of the month in which the product was put into operation or last checked. If, because of the conditions of use and the environment, defects are to be expected earlier, the check must take place before that. Measuring functions are included. For automated external defibrillators for lay users, the STK can be dropped if the device performs self-tests and the operator inspects it visually at regular intervals. For newly commissioned devices, remote monitoring of the self-tests applies in addition from 1 January 2027 (§ 12(2)).

The person doing the check writes a report and marks the product with the year and month of the next check (§ 12(3)). Missing a check is an administrative offense (§ 19 No. 7). The fine can be up to 30,000 euros (§ 94(5) MPDG).

“The safety checks must be carried out at least every two years, by the end of the month in which the product was put into operation or the last safety check was carried out. If, because of the specific conditions of use and the environment of the product, defects are to be expected at an earlier point, the operator is obliged to carry out safety checks, or have them carried out, in good time before the two years expire.”
§ 12(1)MPBetreibV (Medical Devices Operator Ordinance)Original text ↗

Unofficial translation

“From 1 January 2027, for newly commissioned automated external defibrillators intended for use by lay persons, it additionally applies that they must document the results of the self-tests and transmit them regularly and automatically to the operator by remote monitoring.”
§ 12(2)MPBetreibV (Medical Devices Operator Ordinance)Original text ↗

Unofficial translation

What are metrological checks, and which intervals apply?

Metrological checks (MTK) are prescribed for the products in Annex 2, meaning products with a measuring function. They determine whether the product stays within the permitted maximum measurement deviations (§ 15(2)). Proper performance is presumed if the guideline of the national metrology institute, the Physikalisch-Technische Bundesanstalt, is followed (§ 15(1)). It is published in its current version on the institute's website.

The intervals are in Annex 2, in years. Pure-tone and speech audiometers and infrared radiation thermometers: one year. Medical electrical thermometers, thermometers with interchangeable probes, devices for noninvasive blood pressure measurement, eye tonometers and pedal ergometers: two years. Diagnostic dosimeters: five years. For therapy dosimeters, two years are specified, and six years with a control device and semiannual control measurements.

The intervals start at the end of the year in which the product was put into operation or last checked (§ 15(5)). Checks must be done without delay if there are signs that the error limits are exceeded or the metrological properties may have been affected by an intervention. Only standards that are traceable to a national or international standard may be used. After the check, the product must be marked with the year of the next check and the inspecting body (§ 15(7)). Annex 2 was reissued in November 2025, retroactively from 20 February 2025.

“Proper performance of the metrological checks under sentence 1 is presumed if the guideline on metrological checks of products with a measuring function of the Physikalisch-Technische Bundesanstalt is followed.”
§ 15(1)MPBetreibV (Medical Devices Operator Ordinance)Original text ↗

Unofficial translation

“The intervals begin at the end of the year in which the product was put into operation or the last metrological checks were carried out.”
§ 15(5)MPBetreibV (Medical Devices Operator Ordinance)Original text ↗

Unofficial translation

Who may maintain and check medical devices, and what must be checked after maintenance?

Where the ordinance sets special requirements for an activity, it may only be carried out by someone who has current knowledge based on suitable training and relevant professional activity, is not subject to instructions regarding technical assessment and has the means required, including suitable measuring and testing equipment (§ 5). These three conditions apply equally to maintenance, reprocessing, safety and metrological checks and IT security checks. As operator, you verify this at contract signing and whenever the service provider changes staff.

Maintenance includes servicing, inspections and repairs. The manufacturer's instructions must be considered, and appropriate measures must also be taken after safety-relevant incidents (§ 7(2)). After maintenance, the structural and functional features essential for the safety and function of the products must be checked, to the extent the measures could have affected them (§ 7(4)). That applies to every maintenance task and is not limited to the STK.

Anyone who does not carry out maintenance, or does not have it carried out, properly commits an administrative offense (§ 19 No. 2a). That also applies to commissioning an unsuitable provider (§ 19 No. 3). Ask the provider for proof of training and test equipment, and record which devices were checked with which measuring instrument.

“Where special requirements are a prerequisite for an activity under this ordinance, this activity may only be carried out by someone who 1. has current knowledge for the respective activity based on suitable training and relevant professional activity, 2. is not subject to instructions regarding the technical assessment and 3. has the means, in particular rooms, devices and other work equipment, such as suitable measuring and testing equipment, that are required to carry out the respective activity properly and in a traceable manner.”
§ 5MPBetreibV (Medical Devices Operator Ordinance)Original text ↗

Unofficial translation

“After maintenance, the structural and functional features essential for the safety and function of the products must be checked, to the extent they can be affected by the measures.”
§ 7(4)MPBetreibV (Medical Devices Operator Ordinance)Original text ↗

Unofficial translation

How do I keep the inventory register and the medical device logbook, and how long do I retain them?

An inventory register must be kept for all active non-implantable products at each site (§ 14(1)). For each product it contains the designation, type and model, lot code or serial number, year of purchase, the name and address of the manufacturer or authorized representative or importer, the operator's identification number if there is one, and the location and organizational assignment (§ 14(2)). Any data carrier is allowed as long as the data can be made readable within a reasonable time. The register may also be kept within a register that exists under other rules.

For products in Annexes 1 and 2, the medical device logbook is added (§ 13). It contains the identification, the proof of function test and briefing, the names of the person commissioned and of those briefed, the intervals, dates and results of the prescribed safety and metrological checks and IT security checks, the dates of maintenance with the person or company performing it, malfunctions and repeated operating errors, and reports to authorities and manufacturers. The logbook must be accessible to the user during working hours and be kept for five years after the product is taken out of service.

The operator keeps the check reports until the next check. A missing or incomplete logbook is an administrative offense (§ 19 No. 9). vi maintains the inventory register and the deadlines from service reports and invoices and keeps proof ready per device.

“The operator must keep an inventory register for all active non-implantable products at the respective site.”
§ 14(1)MPBetreibV (Medical Devices Operator Ordinance)Original text ↗

Unofficial translation

“intervals and date of performance and the result of prescribed safety and metrological checks and IT security checks, date of maintenance and the name of the person or company that carried out the measure in each case”
§ 13(2) No. 4MPBetreibV (Medical Devices Operator Ordinance)Original text ↗

Unofficial translation

“After the product is taken out of service, the medical device logbook must be kept for five years.”
§ 13(3)MPBetreibV (Medical Devices Operator Ordinance)Original text ↗

Unofficial translation

What applies to software, updates and IT security for networked medical devices?

Maintenance of software includes installing safety-relevant software updates (§ 7(2)). Persons who do not meet the special requirements of § 5 may also perform this installation (§ 7(3) sentence 2). If an update changes how the user handles the software by more than a minor amount, a new briefing is needed after each installation (§ 4(3)). For networked products, the manufacturer's requirements for the digital infrastructure must be observed when connecting to a network (§ 4(6)).

For software that is a medical device of classes IIb and III and for in vitro diagnostic software of classes C and D, § 17 also applies, since 1 August 2025. Before commissioning, the manufacturer or a person authorized by it must check the installation and brief a person commissioned by the operator. The installation check can also take place by remote access. If the software is run in a health care facility, the operator must carry out, or have carried out, appropriate IT security checks at least every two years, by the end of the month of installation or of the last check, and earlier if defects are to be expected. A report with the date and results must be written.

These checks, like all special activities, may only be performed by persons and companies with the right qualifications. If you use remote maintenance, record in the service contract who has access and who approves updates.

“Maintenance of software includes the installation of safety-relevant software updates.”
§ 7(2)MPBetreibV (Medical Devices Operator Ordinance)Original text ↗

Unofficial translation

“The IT security checks must be carried out at least every two years, by the end of the month in which the software was installed or the last IT security check was carried out.”
§ 17(3)MPBetreibV (Medical Devices Operator Ordinance)Original text ↗

Unofficial translation

“For software, this also applies after every installation of software updates that change how the user handles the software during operation and use by more than a minor amount.”
§ 4(3) sentence 2MPBetreibV (Medical Devices Operator Ordinance)Original text ↗

Unofficial translation