viable intelligence.

STK, MTK, inventory list: without gaps.

Clinics, practices and care facilities operate hundreds of medical devices, plus reprocessing, medical gases and building systems with special requirements. The MPBetreibV (German Medical Devices Operator Ordinance) requires checks, inventories and records. vi reads service reports, records and invoices and keeps inventory and deadlines up to date.

Typical assets
  • Defibrillators and patient monitors
  • Infusion and syringe pumps
  • Blood pressure monitors and thermometers
  • Sterilizers and washer-disinfectors
  • Imaging devices
  • Medical gases
  • Ventilators
What this looks like in vi

On the right is an example from this industry: documents arrive in the chat, vi reads them, assigns them and suggests what to do. Assets, companies and amounts are made up.

vi · Example Inc.
Today
Anna Example
When are the next checks due for our medical devices?
07:45
AE
Calendar
vi
Three dates in the first quarter:
DevicesCheckDue
Defibrillators (3)Safety check (STK)Jan 2027
Blood pressure monitors (8)Metrology check (MTK)Feb 2027
Defibrillators (12)Safety check (STK)Mar 2027
07:46
Calendar
vi
For the three defibrillators in January, there is still no appointment with the service provider. Should I request one?
07:46
Message to vi
Typical documents

What lands on your desk.

STK records

Safety checks (STK) for devices under Annex 1 of the MPBetreibV, at least every two years.

MTK records

Metrological checks (MTK) for devices with a measuring function under Annex 2, with error limits according to the guideline of the PTB (Germany's national metrology institute).

Inventory list and medical device logbook

Mandatory records with type, serial number, location and all checks and maintenance.

Service and maintenance contracts

Often one contract per device group, each with its own services, response times and terms.

Validation reports

For reprocessing: sterilizers and washer-disinfectors with validated procedures.

Errors and cost traps

Where things usually go wrong.

01

Device in use without a check

An STK is overdue, yet the device is still in use. The deadline was in the record from the year before last.

02

Inventory list lags behind

Devices are relocated, replaced or decommissioned. The list still shows the old state.

03

Duplicate call-outs

STK and maintenance are ordered separately even though one service provider could do both. Two appointments, two invoices.

04

New ordinance, old lists

A revised MPBetreibV has applied since 20 February 2025. Deadline lists from before that date should be checked against the new version.

What vi does in practice

Read, check, plan.

Inventory from documents

vi builds the inventory list from records, delivery notes and invoices and reconciles it with your list.

STK and MTK in the calendar

Every check gets its deadline, bundled per ward and service provider.

Check records

Failed checks become tickets, and the device is flagged until the proof is in.

Bundle contracts

vi shows where the services of several service providers overlap and where one contract could cover more.

Example

Specialist clinic with three wards

A specialist clinic has several hundred active medical devices, two service providers for STK and MTK and its own biomedical engineering department.

This workflow is an example. Figures and findings are chosen for illustration.

  1. 1

    Upload

    Biomedical engineering uploads last year's records and the existing inventory list as a spreadsheet.

  2. 2

    Reconcile

    vi finds devices with a record that are missing from the list, and entries without proof.

  3. 3

    Plan

    STK and MTK for next year are bundled per ward, with one appointment per service provider and ward.

  4. 4

    Show proof

    For a walkthrough, the list and the records for each device are ready.

Questions

Short answers.

You
Does vi replace biomedical engineering?
vi
No. vi takes the document work off your biomedical engineering team: reading, assigning, keeping deadlines. Checks, assessments and approvals stay with qualified professionals.

Questions in detail.

Expand a question for the answer, the original wording of the rules and links to the official text.

Which devices need a safety inspection, how often, and who may carry it out?

The safety inspection (STK) is required for the products in Annex 1 of the Medical Devices Operator Ordinance (MPBetreibV). These include non-implantable active products such as defibrillators, devices that introduce substances into the bloodstream with potential pressure build-up (for example infusion and syringe pumps), ventilators, magnetic resonance scanners, pressure chambers, hypothermia devices and infant incubators. For other products, and for objects connected to Annex 1 products, the same sentences apply accordingly.

The interval: at the latest every two years, counted to the end of the month in which the product was put into operation or last inspected. If, because of the conditions of use and the environment, defects are to be expected earlier, you must inspect earlier. The safety inspection includes the measuring functions.

Only someone who meets § 5 may carry it out: up-to-date knowledge from training and professional work, no instructions on the technical assessment, and the necessary means, such as suitable measuring and test equipment. The inspector writes a report with the date, results, measured values and measuring methods, and after a successful inspection marks the device with the year and month of the next STK. You keep the report until the next inspection. vi reads the reports and enters the next due date for each device into the calendar.

“For products under Annex 1, the operator must carry out, or have carried out, safety inspections in accordance with the generally recognized rules of technology in line with sentences 2 and 3. The safety inspections must be carried out at the latest every two years, at the end of the month in which the product was put into operation or the last safety inspection was carried out.”
§ 12 para. 1 sentences 1 and 2Medical Devices Operator Ordinance (MPBetreibV)Original text ↗

Unofficial translation

“Where special requirements are prescribed for an activity under this Ordinance, that activity may only be carried out by a person who 1. has up-to-date knowledge for the activity concerned based on suitable training and relevant professional activity, 2. is not subject to instructions regarding the technical assessment, and 3. has the means, in particular premises, devices and other work equipment such as suitable measuring and testing equipment, that are required to carry out the activity concerned properly and in a traceable manner.”
§ 5Medical Devices Operator Ordinance (MPBetreibV)Original text ↗

Unofficial translation

What is a metrological inspection, and which intervals apply?

The metrological inspection (MTK) applies to products with a measuring function listed in Annex 2 of the MPBetreibV. Examples are devices for non-invasive blood pressure measurement, medical thermometers, audiometers, eye pressure tonometers, certain dosimeters and bicycle ergometers. The re-inspection intervals are stated in Annex 2 in years. As examples: two years for non-invasive blood pressure monitors and medical electronic thermometers, one year for infrared radiation thermometers and for audiometers.

The intervals begin at the end of the year in which the product was put into operation or last inspected. Outside the interval, the MTK is required without delay if there are indications that the device does not meet the error limits, or if its metrological properties may have been affected by an intervention or in some other way, for example after a repair. A product in Annex 2 may only be operated if it meets the error limits stated in the guideline of the national metrology institute, the Physikalisch-Technische Bundesanstalt.

You may commission the authorities responsible for metrology, or persons and companies that meet the requirements of § 5. As with the STK, there is a report and a marking with the year of the next inspection. If STK and MTK coincide, the appointments can be combined. vi keeps both intervals for each device in one calendar.

“The periods begin at the end of the year in which the product was put into operation or the last metrological checks were carried out. Metrological checks must be carried out without delay if 1. there are indications that the product does not comply with the error limits under paragraph 2, or 2. the metrological properties of the product may have been affected by an intervention or in some other way.”
§ 15 para. 5 sentences 3 and 4Medical Devices Operator Ordinance (MPBetreibV)Original text ↗

Unofficial translation

“Products of Annex 2 may only be operated or used if they comply with the error limits specified in the guideline under § 15 paragraph 1 sentence 2.”
§ 4 para. 8Medical Devices Operator Ordinance (MPBetreibV)Original text ↗

Unofficial translation

What belongs in the inventory list and the medical device log, and how long do I keep them?

You keep the inventory list for all active non-implantable products at the respective site. For each product, you enter: designation, kind and type, lot code or serial number, year of purchase, name and address of the manufacturer or its authorized representative, or failing that the importer, the internal identification number where there is one, and the location and organizational assignment. Any data carrier is permitted as long as the information can be made legible within a reasonable time.

You keep the medical device log for the products in Annexes 1 and 2, that is, all devices with an STK or MTK. Excluded are electronic clinical thermometers in compact form and blood pressure monitors with a mercury or aneroid gauge. The log holds: identification, proof of the functional test and instruction, the names of the appointed and instructed persons, intervals, date and result of safety and metrological inspections and IT security reviews, date and performer of maintenance, malfunctions and repeated operating errors of the same kind, and details of incident reports. After the device is taken out of service, you keep the log for five years.

The weak point is usually the inventory list: devices are moved, replaced or retired, and the list shows the old state. vi builds the inventory from reports, delivery notes and invoices and compares it with your list.

“The operator must keep an inventory list for all active non-implantable products at the respective site.”
§ 14 para. 1 sentence 1Medical Devices Operator Ordinance (MPBetreibV)Original text ↗

Unofficial translation

“For the products listed in Annexes 1 and 2, the operator must keep a medical device log in accordance with paragraph 2.”
§ 13 para. 1 sentence 1Medical Devices Operator Ordinance (MPBetreibV)Original text ↗

Unofficial translation

“After the product has been taken out of service, the medical device log must be kept for five years.”
§ 13 para. 3 sentence 2Medical Devices Operator Ordinance (MPBetreibV)Original text ↗

Unofficial translation

What do I have to do when a device fails or an incident occurs?

Safety first. Before every use, the user must make sure that the product is functional and in proper condition. If that is in doubt, the device is taken out of service and marked until the matter is clarified. After any maintenance, the design and functional characteristics that are essential for safety and function must be checked, to the extent that the work may have affected them. The service report should show that check.

Second, the report. Anyone who operates or uses products professionally or commercially must report presumed serious incidents to the competent federal authority without delay. An incident already counts if it cannot be ruled out that it is due to a malfunction, a deterioration in performance or a use error, and it led or could have led to death, a serious deterioration in health or a serious threat to public health. Healthcare institutions with regularly more than 20 employees must appoint a medical device safety officer. That person is the contact for authorities and manufacturers and coordinates the internal reporting channels.

Third, the documentation. The medical device log records the date, kind and consequences of malfunctions and the details of reports. vi flags failed inspections and attaches the proof of the repair to the device.

“Anyone who operates or uses products professionally or commercially must report presumed serious incidents that occur in the process to the competent federal authority without delay.”
§ 3 sentence 1Ordinance on the Reporting of Presumed Serious Incidents with Medical Devices (MPAMIV)Original text ↗

Unofficial translation

“After maintenance, the design and functional characteristics that are essential for the safety and functionality of the products must be checked, to the extent that they may have been affected by the measures.”
§ 7 para. 4Medical Devices Operator Ordinance (MPBetreibV)Original text ↗

Unofficial translation

“Healthcare institutions with regularly more than 20 employees must ensure that a competent and reliable person with medical, scientific, nursing, pharmaceutical or technical training is appointed as medical device safety officer.”
§ 6 para. 1 sentence 1Medical Devices Operator Ordinance (MPBetreibV)Original text ↗

Unofficial translation

Who may maintain and repair medical devices, and what belongs in the service contract?

The operator must maintain products and the objects connected to them, or have them maintained. That includes maintenance and inspections that are needed for safe operation and repairs to restore function. The manufacturer's information must be taken into account. Appropriate measures are also needed after safety-relevant incidents. For software, installing safety-relevant updates is part of maintenance.

You may only commission persons, companies or institutions that meet the requirements of § 5 for the product concerned. The installation of safety-relevant software updates is an exception. From that follow the points for the service contract: proof of the qualification of the service staff, a scope of work in line with manufacturer specifications, a functional check after every repair, a service report with date, work, parts replaced, result and software version, response times and a loaner device, and the question of who approves updates and when they are installed.

Also check whether one provider can take on maintenance and inspections together. Then the second trip and the second invoice are no longer needed. Where several providers are responsible for device groups, services often overlap. vi shows such overlaps so you can bundle contracts.

“The operator must maintain products and objects connected to products, or have them maintained.”
§ 7 para. 1 sentence 1Medical Devices Operator Ordinance (MPBetreibV)Original text ↗

Unofficial translation

“Maintenance of software includes the installation of safety-relevant software updates.”
§ 7 para. 2 sentence 4Medical Devices Operator Ordinance (MPBetreibV)Original text ↗

Unofficial translation

What does the operator ordinance require for reprocessing and validation?

Products that are intended to be used in a low-germ or sterile condition must be reprocessed using suitable validated procedures, in such a way that the success is verifiably ensured and nobody is endangered. Single-use devices are excluded. The manufacturer's information must be taken into account. Proper reprocessing is presumed if the joint recommendation of the Commission for Hospital Hygiene and Infection Prevention at the Robert Koch Institute and the Federal Institute for Drugs and Medical Devices on the hygiene requirements for reprocessing is followed.

For products with particularly high requirements, classified in the recommendation as "critical C", certification of the quality management system by a recognized notified body is a prerequisite. You may only commission bodies that meet the requirements of § 5. The validation and performance assessment of the reprocessing process must be carried out on behalf of the operator by qualified specialists, and the competent authority may check this.

For practice, this means: treat validation reports like inspection reports. They belong in the device documentation with date, device, procedure, result and remarks, and you set the next due date according to your quality management requirements. This applies to sterilizers as well as to washer-disinfectors. vi reads validation reports and files due dates and findings for each device.

“The reprocessing of products that are intended to be used in a low-germ or sterile condition, with the exception of single-use devices within the meaning of Article 2 point 8 of Regulation (EU) 2017/745, must be carried out, taking the manufacturer's information into account, using suitable validated procedures in such a way that the success of these procedures is verifiably ensured and the safety and health of patients, users and third parties is not endangered.”
§ 8 para. 1 sentence 1Medical Devices Operator Ordinance (MPBetreibV)Original text ↗

Unofficial translation

“The validation and performance assessment of the reprocessing process must be carried out on behalf of the operator by qualified specialists who meet the requirements of § 5 with regard to the validation and performance assessment of such processes.”
§ 8 para. 4 sentence 3Medical Devices Operator Ordinance (MPBetreibV)Original text ↗

Unofficial translation

What changed with the revised operator ordinance, especially for software and networking?

A revised MPBetreibV has applied since February 2025 (Federal Law Gazette 2025 I No. 38). The current version of Annex 2 with the re-inspection intervals is published in Federal Law Gazette 2025 I No. 263. Interval lists from before that time should therefore be checked against the new version.

Among the changes is networking. When products are connected to a network, the manufacturer's requirements for the digital infrastructure with respect to information security must be observed. Software that is a medical device of classes IIb and III may only be operated if the manufacturer or a person authorized by it has first checked the installation and instructed a person appointed by the operator. In addition come appropriate IT security reviews, at the latest every two years, with a report. They also belong in the medical device log.

For automated external defibrillators intended for laypersons, the STK can be dropped if the device is self-testing and the operator carries out regular visual checks. For newly commissioned devices, from January 1, 2027 they must also document self-tests and transmit the results automatically to the operator on a regular basis. Keep track of which devices in your inventory are affected. vi keeps software versions and inspection records together for each device.

“For networked products, when connecting to a network, the manufacturer's requirements regarding the digital infrastructure in relation to the information security of its products must be observed.”
§ 4 para. 6 sentence 3Medical Devices Operator Ordinance (MPBetreibV)Original text ↗

Unofficial translation

“The IT security reviews must be carried out at the latest every two years, at the end of the month in which the software was installed or the last IT security review was carried out.”
§ 17 para. 3 sentence 2Medical Devices Operator Ordinance (MPBetreibV)Original text ↗

Unofficial translation

“From January 1, 2027, for newly commissioned automated external defibrillators intended for use by laypersons, it additionally applies that they document the results of the self-tests and regularly transmit them automatically to the operator by remote monitoring.”
§ 12 para. 2 sentence 2Medical Devices Operator Ordinance (MPBetreibV)Original text ↗

Unofficial translation

Maintenance that thinks ahead.

Show us a service contract or inspection report. We'll show you what vi makes of it.